Pulse CO Oximeters Not Accurate for Assessing Blood Carboxyhemoglobin
THURSDAY, Oct. 20, 2022 (HealthDay News) -- The use of noninvasive pulse carbon monoxide (CO) oximeters (SpCO) is not accurate for determining blood carboxyhemoglobin (COHb) levels, according to a study presented Oct. 18 at the European Emergency Medicine Congress, in Germany.
Mathilde Papin, from Nantes University Hospital in France, and colleagues conducted a meta-analysis to determine the accuracy of SpCO in determining blood COHb level. The systematic review included 19 studies; 11 and 13 were included for quantitative analysis of the primary and secondary end points, respectively.
Three different devices were used to measure SpCO. The principal source of bias and concern regarding applicability was patient selection. There was no publication bias. The researchers found that the area under the summary receiver operating characteristic curve was 86 percent, with pooled sensitivity and specificity of 0.77 and 0.83, respectively. The mean bias was 0.95 percent, and limits of agreement were −6.08 to 8.00 percent.
"At 23 percent, the false negative rate with pulse oximetry is too high for reliably triaging patients with suspected carbon monoxide poisoning. This method is not accurate enough and should not be used in clinical practice," the authors write.
Related Posts
Una aspirina de dosis baja al día sube las probabilidades de sangrado cerebral, sin bajar el riesgo de ACV
JUEVES, 27 de julio de 2023 (HealthDay News) -- Durante años, los adultos...
Biden utiliza una máquina de PPCVR de noche para ayudar a aliviar la apnea del sueño
JUEVES, 29 de junio de 2023 (HealthDay News) -- El Presidente Joe Biden comenzó...
ADHD Med Exposure in Utero Not Tied to Long-Term Impact in Offspring
FRIDAY, Feb. 24, 2023 (HealthDay News) -- Neurodevelopment and growth in...
One Key to Good Sleep for Teens: No Social Media at Bedtime
TUESDAY, May 30, 2023 (HealthDay News) -- Teens need their sleep, and a new...